UJI KELAYAKAN APRON DENGAN MENGGUNAKAN IMAGING PLATE (IP) DI INSTALASI RADIOLOGI RUMAH SAKIT YASMIN BANYUWANGI

  • Sri Sugiarti STIKes WCH
Keywords: Apron, Imaging Plate, Leak Test

Abstract

Test theĀ  Feasibility of the Apron Using an Imaging Plate (IP) at the Radiology Installation Yasmin Banyuwangi Hospital. Apron leak test at the Radiology Installation Yasmin Banyuwangi Hospital using an imaging plate on X-ray aircraft without fluoroscopy. The purpose of this study is to determine the number of aprons that are suitable for use in the Radiology Installation of Yasmin Banyuwangi Hospital. The research was carried out by physical apron testing, then irradiate the surface of the apron using X-ray aircraft without fluoroscopy equipped by dividing 4 parts on a cassette imaging plate. The results of radiographs from three aprons after being interpreted by 3 respondents indicated that been tested in a Radiology Installation of Yasmin Banyuwangi Hospital only number three apron was feasible, although the physical condition of the apron is not good and there is no fracture on the results of the radiograph. While the number one and two aprons are not feasible to use because the physical condition of the apron is not good and some exceed the eligibility criteria used in the radiograph.

References

Akhadi, 2000. Dasar-Dasar Proteksi Radiasi.Jakarta : Rineka Cipta.
Roser, 2010. Quality Asurance of X-Ray Protection Clothing at the University Hospital Basel. . Unervesity Hospital Basel: Radiological Physic.
Keputusan Menteri Kesehatan RI No.1250/MENKES/SK/XII/2009.
Nasir, 2011. Metode Penelitian. Bogor: Ghalia Indonesia.
Kasiram, 2010. Metodologi Penelitian. Malang : UIN Maliki Press.
Sugiyono, 2010. Statistik Untuk Penelitian. Bandung : Alfabeta.
Usman, 2008. Metode Penelitian Sosial. Jakarta : Bumi Aksara.
Oyar, 2011. How Protective Are The Lead Apron We Use Against Ionizing Radiation?.(On Line), (https://www.researchgate.net/publication/,diakses 12 Maret 2019)
Peraturan Kepala Badan Pengawas Tenaga Nuklir No. 8 Tahun 2011 tentang Keselamatan Radiasi Dalam Penggunaan Pesawat Sinar-x Radiologi Diagnostik dan Intervensional, 2011. Jakarta: PT. Armas Duta Jaya.
Grover, 2002. Regulatory Bodies, Safety Norms, Does Limits And Protection Devices. USA : Cengange Learning.
Published
2021-06-28
How to Cite
Sugiarti, S. (2021). UJI KELAYAKAN APRON DENGAN MENGGUNAKAN IMAGING PLATE (IP) DI INSTALASI RADIOLOGI RUMAH SAKIT YASMIN BANYUWANGI. Health Care Media, 5(1), 8 - 15. Retrieved from https://stikeswch-malang.e-journal.id/Health/article/view/158
Section
Healt Care Media